Extech FL700 Fluoride Meter User Manual

Загвар: FL700

1. Танилцуулга

The Extech FL700 Fluoride Meter is a precision instrument designed for the rapid and accurate measurement of fluoride ions in various aqueous samples, including drinking water. This device integrates a sensing electrode, measurement electronics, and a digital display into a single, portable unit, making it suitable for both field and laboratory applications.

The FL700 provides direct readings of total fluoride concentrations ranging from 9.99 ppm or mg/L down to 0.1 ppm. It features automatic temperature compensation to ensure measurement accuracy across varying environmental conditions.

2. Аюулгүй байдлын мэдээлэл

Please read and understand all safety precautions before operating the Extech FL700 Fluoride Meter. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always handle the fluoride electrode with care. Avoid touching the sensing element directly.
  • When working with TISAB reagent tablets or any chemical solutions, wear appropriate personal protective equipment, including gloves and eye protection.
  • Do not immerse the entire meter in liquid. The meter is designed for electrode immersion only.
  • Store the meter and reagents in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Хүүхдийн гар хүрэхгүй газар хадгална.

3. Бүтээгдэхүүн дууссанview

The Extech FL700 Fluoride Meter is a compact and user-friendly device. Key components and features are outlined below:

Extech FL700 Fluoride Meter front view
Урд view of the Extech FL700 Fluoride Meter, showing the display and control buttons.
Hand holding Extech FL700 Fluoride Meter
The Extech FL700 Fluoride Meter held in hand, demonstrating its ergonomic design and clear display.
Close-up of Extech FL700 electrode tip
Ойролцоогоор view of the fluoride electrode tip, which is integrated into the meter.
Extech FL700 Fluoride Meter in use with sample
The Extech FL700 Fluoride Meter being used to measure a liquid sample, showing the meter in its stand.
Extech FL700 Fluoride Meter and included accessories
All components included with the Extech FL700: the meter, sensor cap, TISAB reagent tablets, and neckstrap.

Гол онцлогууд:

  • Хоёр цэгийн шалгалт тохируулга: Ensures accurate measurements across the fluoride range.
  • Санах ойн сан: Stores up to 25 labeled readings for later review.
  • Field or lab testing: Designed for versatile use in various environments.
  • Хурдан хариу үйлдэл: Provides fluoride measurements in less than 1 minute.
  • Large dual LCD: Simultaneously displays ppm (or mg/L) and temperature with a bargraph.

4. Тохиргоо

4.1. Зай суурилуулах

The Extech FL700 requires four 3V button batteries (included). To install or replace batteries:

  1. Twist off the battery compartment cap located at the top of the meter.
  2. Insert the four 3V button batteries, ensuring correct polarity as indicated inside the compartment.
  3. Replace the battery compartment cap and twist to secure.

4.2. Анхны бэлтгэл

Before first use or after prolonged storage, prepare the electrode:

  1. Remove the sensor cap from the electrode tip.
  2. Rinse the electrode tip with distilled water.
  3. Condition the electrode by immersing it in a fluoride standard solution (e.g., 1 ppm or 10 ppm) for approximately 15-30 minutes.

5. Ашиглалтын заавар

5.1. Асаах/унтраах

дээр дарна уу ON/OFF тоолуурыг асаах эсвэл унтраах товчлуур.

5.2. Calibration (Two-Point)

Accurate calibration is crucial for reliable measurements. The FL700 supports two-point calibration.

  1. Prepare two fluoride standard solutions, typically one low concentration (e.g., 1 ppm) and one high concentration (e.g., 10 ppm). Ensure the solutions are at a stable temperature.
  2. Электродыг нэрмэл усаар зайлж, хатаана.
  3. Immerse the electrode into the first standard solution.
  4. -г удаан дар CAL / RECALL button until 'CAL' appears on the display. The meter will automatically recognize the standard and calibrate. Wait for the reading to stabilize.
  5. Rinse the electrode again and immerse it into the second standard solution.
  6. -г удаан дар CAL / RECALL button again. The meter will complete the two-point calibration.
  7. The meter is now calibrated and ready for measurement.

5.3. Хэмжилт хийх

To measure fluoride concentration in a sampле:

  1. Тоолуурыг тохируулсан эсэхийг шалгана уу.
  2. Электродыг нэрмэл усаар зайлж, хатаана.
  3. Add TISAB (Total Ionic Strength Adjustment Buffer) reagent to your sample according to the reagent instructions. TISAB ensures that the fluoride ions are in a measurable form and minimizes interference.
  4. Immerse the electrode tip into the prepared sample solution. Ensure the tip is fully submerged.
  5. Allow the reading to stabilize on the display. The large dual LCD will show the fluoride concentration in ppm (or mg/L) and the temperature.
  6. Одоогийн уншилтыг хадгалахын тулд товчийг дарна уу MODE / HOLD товчийг дарна уу. Суллахын тулд дахин дарна уу.

5.4. Recalling Stored Readings

The meter can store up to 25 labeled readings.

  • To recall stored readings, press the CAL / RECALL button briefly when not in calibration mode.
  • -г ашиглана уу MODE / HOLD button to scroll through the stored readings.

6. Засвар үйлчилгээ

6.1. Электрод цэвэрлэх

Regular cleaning of the electrode ensures optimal performance.

  • After each use, rinse the electrode thoroughly with distilled water.
  • If the electrode becomes fouled or readings become erratic, clean it with a mild detergent solution or a specialized electrode cleaning solution. Rinse thoroughly with distilled water afterward.

6.2. Хадгалах

Proper storage extends the life of the electrode and meter.

  • Always replace the sensor cap on the electrode tip when not in use. The cap should contain a small amount of electrode storage solution or a fluoride standard solution to keep the membrane hydrated. Do not store dry.
  • Тоолуурыг сэрүүн, хуурай газар хадгална.

6.3. Зай солих

Replace batteries when the low battery indicator appears on the display or when the meter fails to power on. Refer to Section 4.1 for battery installation instructions.

7. Алдаа засах

АсуудалБоломжит шалтгаанШийдэл
Тоолуур асахгүй байнаДууссан эсвэл буруу суурилуулсан батерейБатерейны туйлшралыг шалгах; батерейг солих;
Тогтворгүй эсвэл тогтворгүй заалтуудDirty electrode; air bubbles on electrode; insufficient sample volume; expired or incorrect TISAB.Clean electrode; gently tap electrode to remove bubbles; ensure electrode is fully submerged; check TISAB.
Meter fails to calibrateIncorrect standard solutions; dirty electrode; electrode not conditioned; damaged electrode.Verify standard concentrations; clean and condition electrode; replace electrode if damaged.
Батарей бага үзүүлэлтБатерей бага байна.Replace all four 3V button batteries.

8. Техникийн үзүүлэлтүүд

ТодорхойлолтДэлгэрэнгүй
ҮйлдвэрлэгчExtech
Хэсгийн дугаарFL700
Зүйлийн жин7.7 унц
Бүтээгдэхүүний хэмжээс6.8 x 1.6 x 1.4 инч
Загварын дугаарFL700
Батерей4 Лити Металл батерей шаардлагатай (дагсан)
ЗагварFL700
Барааны багцын тоо1
Хэсгийн тоо1
Оруулсан бүрэлдэхүүн хэсгүүдНэгж, зааварчилгаа
Зайны эсийн төрөлШүлтлэг
Эхний боломжтой огноо15 оны долдугаар сарын 2008

9. Баталгаат хугацааны мэдээлэл

The Extech FL700 Fluoride Meter comes with a full one-year warranty on parts and workmanship, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modification.

10. Дэмжлэг

For technical support, service, or additional information, please visit the official Extech webсайт эсвэл харилцагчийн үйлчилгээний хэлтэстэй холбогдоно уу. Та мөн илүү дэлгэрэнгүй мэдээлэл, бүтээгдэхүүнийг эндээс авах боломжтой Extech Store on Amazon.

Холбогдох баримт бичиг - FL700

Урьдчилсанview EXTECH 45170 4-ин-1 чийгшил, температур, агаарын урсгал, гэрлийн тоолуурын хэрэглэгчийн гарын авлага
EXTECH 45170-ийн хэрэглэгчийн гарын авлага, чийгшил, температур, агаарын урсгал, гэрлийг хэмжих 4-ийн 1 метр. Ажиллагаа, техникийн үзүүлэлтүүд болон баталгаат байдлын мэдээлэл багтана.
Урьдчилсанview Radiazione Utente Misuratore гарын авлага UV-AB Extech UV505
Radiazione UV-AB Extech UV505-ийн Misuratore-ийн бүрэн гарын авлага. Ажиллагааны ажил, тусгай техник, мэдээлэл, мэдээлэл зэргийг оруулаарай.
Урьдчилсанview Extech UV505 UV-AB гэрлийн тоолуурын хэрэглэгчийн гарын авлага
Extech UV505 UV-AB гэрлийн тоолуурын хэрэглэгчийн гарын авлага. Энэхүү баримт бичигт төхөөрөмжийн онцлог, ажиллагаа, техникийн үзүүлэлтүүд, аюулгүй байдлын урьдчилан сэргийлэх арга хэмжээ, хэт ягаан туяаны болон хэт ягаан туяаны гэрлийг хэмжих засвар үйлчилгээний талаарх дэлгэрэнгүй мэдээллийг агуулсан болно.
Урьдчилсанview Extech EX810 1000 Amp Klemmeter уулзсан IR термометр Gebruikershandleiding
Extech EX810 1000-д зориулсан нэмэлт төхөөрөмж Amp klemmeter met inebouwde IR термометр. Зааварчилгаа, техникийн үзүүлэлтүүд болон техникийн үзүүлэлтүүд, угаалгын өрөө, угаалгын өрөө, багтаамж, температур зэргийг багтаасан болно.
Урьдчилсанview Extech MO257 Pinless Moisture Meter User Guide
User guide for the Extech MO257 Pinless Moisture Meter, detailing its features, operation, specifications, and maintenance for accurate non-invasive moisture measurements in building materials.
Урьдчилсанview Extech EX650 цуврал хэрэглэгчийн гарын авлага: 600A True RMS Digital Clamp Метр
Extech EX650 Series 600A True RMS дижитал cl-ийн хэрэглэгчийн дэлгэрэнгүй гарын авлага.amp метр, EX650 ба EX655 загваруудыг хамарсан. Тохиромжтой хувьсах гүйдлийн AC/DC-ийн үйл ажиллагаа, онцлог, аюулгүй байдал, техникийн үзүүлэлтүүдийн талаар мэдэж аваарайtage, гүйдэл, эсэргүүцэл ба бусад хэмжилтүүд.