The GIMA 28211 Super Vega Suction Aspirator is a compact and reliable medical device designed for the aspiration of bodily fluids such as mucus, catarrh, or blood. It is suitable for nasal, oral, and tracheal aspiration in both adults and children. This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your device.

GIMA 28211 Super Vega Suction Aspirator main unit

Image 1.1: The GIMA 28211 Super Vega Suction Aspirator, showing the main unit with collection jar, tubing, and control panel.

2. Аюулгүй байдлын мэдээлэл

  • Read this manual completely before operating the device.
  • Ашиглалтын явцад төхөөрөмжийг тогтвортой, тэгш гадаргуу дээр байрлуулсан эсэхийг шалгаарай.
  • Үндсэн төхөөрөмжийг ус болон бусад шингэнд дүрж болохгүй.
  • Only use accessories and spare parts approved by GIMA.
  • Regularly check the power cord for damage. Do not use if damaged.
  • Төхөөрөмжийг хүүхэд болон зөвшөөрөлгүй хүмүүсээс хол байлга.
  • This device is intended for medical use and should be operated by trained individuals or under medical supervision.
  • Ашиглалтын явцад төхөөрөмжийн эргэн тойронд агааржуулалт сайтай байх ёстой.

3. Багцын агуулга

Савлахдаа бүх бүрэлдэхүүн хэсгүүд байгаа бөгөөд гэмтэлгүй эсэхийг шалгана уу:

  • GIMA 28211 Super Vega Suction Aspirator main unit
  • 1000 ml autoclavable collection jar with lid
  • Suction tubing set
  • Бактерийн шүүлтүүр
  • Цахилгааны утас
  • Хэрэглэгчийн гарын авлага
Collection jar and lid

Image 3.1: 1000 ml collection jar with lid.

Бактерийн шүүлтүүр

Image 3.2: Bacterial filter.

Цахилгааны утас

Image 3.3: Power cord.

4. Тохиргоо

  1. Байршил: Place the aspirator on a stable, flat, and dry surface, ensuring adequate space for ventilation.
  2. Connect Collection Jar: Securely attach the 1000 ml collection jar to the designated holder on the side of the main unit. Ensure the lid is tightly sealed.
  3. Хоолойг холбох:
    • Connect one end of the short silicone tube to the 'VACUUM' port on the collection jar lid.
    • Connect the other end of this short tube to the suction inlet on the main unit (usually marked 'SUCTION' or similar).
    • Connect one end of the longer silicone tube to the 'PATIENT' port on the collection jar lid. This end will connect to the patient catheter.
  4. Install Bacterial Filter: Insert the bacterial filter between the 'VACUUM' port on the collection jar lid and the suction inlet of the main unit. This prevents fluid overflow and protects the pump. Ensure the filter is correctly oriented according to any arrows on its casing.
  5. Эрчим хүчний холболт: Connect the power cord to the aspirator's power inlet and then plug it into a suitable electrical outlet (220-230 V/50-60 Hz).
GIMA 28211 Super Vega Suction Aspirator fully assembled

Image 4.1: The aspirator fully assembled and ready for operation.

5. Ашиглалтын заавар

  1. Асаах: Press the green power switch (marked with 'I/O') located on the front panel to turn on the device. The indicator light should illuminate.
  2. Соролтыг тохируулах: Use the 'SUCTION CONTROL' knob on the front panel to adjust the desired vacuum level. Turn clockwise to increase suction and counter-clockwise to decrease it.
  3. Тоос сорогч хянах: Observe the vacuum gauge on the front panel to ensure the desired suction pressure is achieved. The gauge displays pressure in mmHg and kPa.
  4. Катетерийг холбох: Attach the appropriate sterile suction catheter to the free end of the patient tubing.
  5. Perform Aspiration: Carefully perform the aspiration procedure as directed by a medical professional.
  6. Унтраах: After use, press the power switch to turn off the device.
Control panel of GIMA 28211 Super Vega Suction Aspirator

Image 5.1: Close-up of the control panel, showing the power switch, vacuum gauge, and suction control knob.

6. Засвар үйлчилгээ, цэвэрлэгээ

Regular cleaning and maintenance are crucial for the device's longevity and hygienic operation.

6.1. Collection Jar

  • Empty the collection jar after each use or when the fluid level approaches the maximum fill line.
  • The 1000 ml collection jar is autoclavable at 121°C. An optional jar is available that can be autoclaved at 134°C.
  • Wash the jar and lid thoroughly with warm water and a mild detergent. Rinse well and sterilize as appropriate.
  • Ensure the safety float valve in the lid is clean and moves freely to prevent overflow.

6.2. Tubing and Filter

  • Clean the suction tubing after each use by aspirating clean water through it.
  • Replace the bacterial filter regularly, especially if it becomes wet, discolored, or shows signs of contamination. The filter is disposable.

6.3. Үндсэн хэсэг

  • Wipe the exterior of the main unit with a soft cloth dampened with a mild disinfectant.
  • Зүлгүүрийн цэвэрлэгч, уусгагчийг бүү ашигла.
  • Ensure no liquids enter the internal components of the device.

7. Алдаа засах

АсуудалБоломжит шалтгаанШийдэл
Төхөөрөмж асахгүй байнаNo power supply; damaged power cord; faulty power switchCheck power cord connection and wall outlet. Inspect power cord for damage. If problem persists, contact technical support.
No or weak suctionLoose connections; clogged tubing/catheter; full collection jar; faulty bacterial filter; pump issueCheck all tubing connections for leaks. Clear any blockages in tubing/catheter. Empty collection jar. Replace bacterial filter. If problem persists, contact technical support.
Fluid overflow into main unitCollection jar overfilled; faulty safety float valve; bacterial filter not installed or cloggedEmpty collection jar immediately. Check float valve for proper movement. Ensure bacterial filter is correctly installed and not clogged.
Ашиглалтын явцад ер бусын дуу чимээLoose components; internal pump issueEnsure all external components are secure. If noise persists, discontinue use and contact technical support.

8. Техникийн үзүүлэлтүүд

  • Загвар: GIMA 28211 Super Vega
  • Үйл ажиллагааны ботьtage: 220-230 В / 50-60 Гц
  • Эрчим хүчний хэрэглээ: 106 Вт
  • Collection Jar Capacity: 1000 мл
  • Хэмжээ (L x W x H): 35 x 21 x 18 см
  • Жин: Ойролцоогоор 3.34 кг
  • Вакуум зохицуулагч: Тасралтгүй тохируулж болно
  • Autoclavable Jar: 121°C (optional 134°C)
  • Тусгаарлагч: Electrical and thermal insulating plastic body

9. Баталгаат хугацаа ба дэмжлэг

Information regarding the specific warranty period for the GIMA 28211 Super Vega Suction Aspirator is not provided in this document. Please refer to your purchase documentation or contact your local GIMA distributor or customer service for warranty details and technical support.

For assistance, please have your model number (28211) and purchase date available.

Холбогдох баримт бичиг - 28211

Урьдчилсанview GIMA SUPERVEGA BATTERY SUCTION UNIT User Manual and Technical Guide
This document provides comprehensive operating instructions, safety guidelines, technical specifications, and maintenance procedures for the GIMA SUPERVEGA BATTERY SUCTION UNIT (Model 28193), a professional medical suction device.
Урьдчилсанview GIMA TOBI MANUALE Сорогч аспиратор - Хэрэглэгчийн гарын авлага ба техникийн үзүүлэлтүүд
GIMA TOBI MANUALE сорох аспираторын хэрэглэгчийн дэлгэрэнгүй гарын авлага болон техникийн үзүүлэлтүүд. Үүний онцлог, аюулгүй байдлын заавар, цэвэрлэгээ, засвар үйлчилгээ, алдааг олж засварлах талаар олж мэдээрэй.
Урьдчилсанview GIMA S/S Bierer Aspiration Tube - User Manual & Specifications (Models 28262, 28264)
Official user manual and technical specifications for the GIMA S/S Bierer Aspiration Tube (Models 28262 and 28264). Details intended use, features, procedure, cleaning, sterilization, operating conditions, and manufacturer information.
Урьдчилсанview GIMA импульсийн оксиметрийн датчикийн хэрэглэгчийн гарын авлага - Техникийн үзүүлэлт ба хэрэглээний гарын авлага
GIMA импульсийн оксиметрийн зондуудын цогц хэрэглэгчийн гарын авлагад бүтээгдэхүүний үзүүлэлт, зориулалт, клиник ашиг тус, засвар үйлчилгээ, цэвэрлэгээ, ариутгал, хүрээлэн буй орчны шаардлага, цахилгаан соронзон нийцтэй байдал (EMC)-ийн мэдээллийг багтаасан болно. ESA0015, ESC0029 загварын дугаарууд болон бусад мэдээллийг багтаасан болно.
Урьдчилсанview GIMA KYOLING® шүүлтүүртэй гар сорох насос - Бүтээгдэхүүний мэдээлэл
Шүүлтүүртэй GIMA KYOLING® гарын авлагын сорох насосны дэлгэрэнгүй гарын авлага, түүний онцлог, техникийн үзүүлэлт, ашиглалтын заавар, засвар үйлчилгээний журам, эмнэлгийн мэргэжилтнүүдэд зориулсан алдааг олж засварлах зөвлөмжүүд.
Урьдчилсанview GIMA даралтат идэвхжүүлсэн хамгаалалтын ланцет (PA) - Капилляр цусны S-д зориулсан эмнэлгийн төхөөрөмжampлинг
GIMA даралтат идэвхижүүлсэн хамгаалалтын ланцетуудын (Загвар PA) тухай мэдээлэл, хялгасан судасны цусыг нэг удаагийн ариутгасан төхөөрөмжampлинг. Зориулалт, заавар, сэрэмжлүүлэг, хадгалах, устгах зэргийг багтаасан болно.