INSTRUKART TU-2016

INSTRUKART Lutron TU-2016 булингар хэмжигч хэрэглэгчийн гарын авлага

Model: TU-2016

1. Танилцуулга

This manual provides detailed instructions for the proper operation, maintenance, and troubleshooting of the INSTRUKART Lutron TU-2016 Turbidity Meter. Please read this manual thoroughly before using the device to ensure accurate measurements and safe operation.

The Lutron TU-2016 is designed for measuring turbidity in various water sources, including lakes, water tanks, rivers, and mangroves. It features a wide measurement range, high resolution, and a user-friendly interface, adhering to ISO 7027 standards.

2. Аюулгүй байдлын мэдээлэл

3. Багцын агуулга

Савыг задлахдаа доор жагсаасан бүх зүйл байгаа бөгөөд сайн нөхцөлд байгаа эсэхийг шалгаарай.

INSTRUKART Lutron TU-2016 Turbidity Meter and accessories in a hard carrying case, including standard solutions, an empty sample bottle, and the instruction manual.

Зураг 3.1: Contents of the TU-2016 Turbidity Meter package, neatly organized within its hard carrying case.

INSTRUKART Lutron TU-2016 Turbidity Meter displayed with various accessories including standard solution bottles, an empty sample bottle, and a cleaning cloth.

Зураг 3.2: The TU-2016 Turbidity Meter along with its essential accessories for measurement and maintenance.

4. Бүтээгдэхүүн дууссанview

The Lutron TU-2016 Turbidity Meter is a portable, microprocessor-controlled device designed for precise turbidity measurements. Key features include:

Урд view of the INSTRUKART Lutron TU-2016 Turbidity Meter, showing its LCD display with a reading of 11.45 NTU and control buttons.

Зураг 4.1: Урд view of the TU-2016 Turbidity Meter, highlighting the display and control panel.

A hand holding the INSTRUKART Lutron TU-2016 Turbidity Meter, demonstrating its ergonomic design and portable size. The display shows 0.0 NTU.

Зураг 4.2: The TU-2016 Turbidity Meter held in hand, illustrating its compact and portable design.

5. Тохиргоо

5.1 Батерей суурилуулах

The meter operates on six DC 1.5 V AAA (UM4) batteries. To install or replace batteries:

  1. Тоолуурын арын хэсэгт зайны тасалгааны тагийг олоорой.
  2. Open the cover and insert six AAA batteries, observing the correct polarity (+/-).
  3. Зайны тасалгааны тагийг сайтар хаа.

5.2 Анхдагч асаалттай

дээр дарна уу ХҮЧ button to turn on the meter. The LCD display will illuminate and show a brief self-test before entering measurement mode.

6. Ашиглалтын заавар

6.1 Preparing a Sample

  1. Ensure the empty testing bottle (cuvette) is clean and dry. Wipe the exterior with a lint-free cloth.
  2. Fill the testing bottle with the water sample to be measured. Avoid air bubbles.
  3. Securely cap the bottle.

6.2 Хэмжилт хийх

  1. s-г нээнэ үүample compartment lid on the top of the meter.
  2. A hand opening the sample compartment lid of the INSTRUKART Lutron TU-2016 Turbidity Meter.

    Зураг 6.1: Opening the sampтасалгааны таг.

  3. Insert the prepared sample bottle into the compartment, ensuring it is seated correctly.
  4. A hand inserting a sample bottle into sample compartment of the INSTRUKART Lutron TU-2016 Turbidity Meter.

    Зураг 6.2: Оруулж байнаample bottle into the meter.

  5. s-г хааampтасалгааны таг.
  6. The INSTRUKART Lutron TU-2016 Turbidity Meter with its sample compartment lid open, next to an empty sampле лонх.

    Зураг 6.3: The meter with the sample compartment open, ready for sample insertion.

  7. дээр дарна уу TEST/CAL button to initiate a measurement. The turbidity reading will appear on the LCD display.

6.3 Өгөгдөл хадгалах функц

Дэлгэц дээрх одоогийн заалтыг хөлдөөхийн тулд товчийг дарна уу БАРЬ button. Press it again to release and resume live measurement.

6.4 Memory Recall (MAX/MIN)

The meter can record and recall the maximum and minimum turbidity readings:

6.5 Автомат унтрах

The meter features an auto shut-off function to conserve battery life. If no button is pressed for a certain period, the meter will automatically power off. This can be manually overridden by pressing the ХҮЧ төхөөрөмжийг унтраах товчийг дарна уу.

7. Шалгалт тохируулга

The TU-2016 features a three-point push-button calibration system. Regular calibration ensures the accuracy of your measurements. Use the provided 0 NTU and 10 NTU standard solutions for calibration.

  1. Тоолуурыг асаана уу.
  2. Insert the 0 NTU standard solution into the sampтасалгаа.
  3. -г удаан дар TEST/CAL button until the calibration mode is indicated on the display. Follow the on-screen prompts to calibrate the 0 NTU point.
  4. Remove the 0 NTU solution and insert the 10 NTU standard solution.
  5. Repeat the calibration process for the 10 NTU point.
  6. The third calibration point (if applicable) can be performed similarly with a higher NTU standard solution if available and required for your specific application.
  7. After successful calibration, the meter will return to measurement mode.

Note: Refer to the Factory Calibration Certificate for details on the initial calibration performed by the manufacturer.

8. Засвар үйлчилгээ

8.1 Cleaning the Meter

8.2 Хадгалалт

9. Алдаа засах

АсуудалБоломжит шалтгаанШийдэл
Тоолуур асахгүй байна.Дууссан эсвэл буруу суурилуулсан батерей.Батерейг солих, зөв ​​туйлшралыг хангах.
Буруу уншилтууд.
  • Бохир сampле лонх.
  • s дахь агаарын бөмбөлөгүүдample.
  • Тоолуур нь тохируулга шаарддаг.
  • Expired standard solutions.
  • Цэвэр сample bottle thoroughly.
  • Gently tap bottle to remove bubbles.
  • Perform a 3-point calibration.
  • Use fresh standard solutions.
Дэлгэц дээр "Err" эсвэл үүнтэй төстэй алдаа гарч ирнэ.Measurement range exceeded or internal error.s-г баталгаажуулна ууample is within 0-1000 NTU range. Power cycle the device. If error persists, contact support.

10. Техникийн үзүүлэлтүүд

ПараметрДэлгэрэнгүй
ХэлхээCustom one-chip of microprocessor LSI circuit
ДэлгэцLCD хэмжээ: 41 мм x 34 мм
Хэмжилтийн хүрээ0.00 to 50.00 NTU, 50 to 1,000 NTU (Auto range)
Шийдвэр0.01 NTU (0.00 to 50.00 NTU), 1 NTU (50 to 1,000 NTU)
Нарийвчлал± 5 % F.S. or ± 0.5 NTU, whichever is greater
Гэрлийн эх үүсвэрLED, 850 нм
ИлрүүлэгчФото диод
Стандарт нийцэлMeet ISO 7027
Хариу өгөх хугацаа10 секундээс бага
Sample Volume Needed10 мл
Өгөгдөл хадгалахДэлгэцийн уншилтыг царцаах
Санах ойг эргэн санахХамгийн их ба хамгийн бага утга
Дэлгэц Sampцаг хугацааОйролцоогоор 1 секунд
УнтраахТовчлуураар автомат унтрах эсвэл гараар унтраах
Ашиглалтын температур0-аас 50 хэм хүртэл
Ашиглалтын чийгшил85% -иас бага RH
Эрчим хүчний хангамжDC 1.5 V battery (UM4, AAA) x 6 PCs
Power Current (Standby)Ойролцоогоор. DC 3.5 мА
Power Current (Testing)Ойролцоогоор. DC 36 мА
Жин320 g Approx. (0.71 lbs)
Хэмжээ155 x 76 x 62 мм (6.1 x 2.99 x 2.44 инч)

11. Дэмжлэг ба баталгаа

For technical assistance, troubleshooting beyond this manual, or warranty inquiries, please contact INSTRUKART customer support. Refer to your purchase documentation for specific warranty terms and contact information.

The product comes with a Factory Calibration Certificate, ensuring its initial accuracy and compliance with international standards.